|
| Time
Resolved Optical (TRO) Spectroscopy |
|
|
|
|
|
|
| Photomultiplier tube (PMT) detector (single frequency detection) |
|
|
|
|
|
|
 |
A
diagram of the TRO system is shown below. The pump source is a NY-61 Nd:YAG
laser manufactured by Continuum. Power is attenuated to approximately 20
mJ/pulse for the third harmonic (355 nm). The repetition rate is programmable
from a single shot to 2, 4, 5, 10 and 20 Hz with 2 Hz being the most commonly
used for these experiments. A 300 W xenon arc lamp from ILC Technology supplied
the light for the probe. For kinetic traces, the output of the probe is
first passed through an IR filter (water) and then through a SPEX single
monochromator with the slits wide open. A band pass filter is substituted
for the single monochromator for some of the studies in order to facilitate
switching to CCD detection. A shutter is placed after the monochromator
so that the probe beam could be shut off when data was not being taken.
This configuration ensured that probe induced photochemistry will be kept
to a minimum. The probe is focused into the sample and finally through a
SPEX Model 1680 double grating monochromator and onto a RCA IP28 photomultiplier
tube. The temporal response is recorded on a Tektronix TDS 540 digital oscilloscope
which is linked to a 50 MHz 486 PC. Transients are commonly recorded as
100 shot averages with the best results being observed when the pump and
probe beams are collinear to each other instead of perpendicular. Alignment
of the system is checked prior to each experiment by optimizing the signal
of ZnTPP at 450 nm. The optical arrangement allows one to very quickly change
from using a PMT detector to a CCD detector. This is accomplished by focusing
into a fiber optic which is placed in front of the double monochromator.
|
|
|
 |
 |
|
| Charge
Coupled Device (CCD) Detector |
|
| Similarly
the spectral response is obtained by passing the probe beam (all wavelengths)
through the sample then focusing into a fiber optic coupled to a SpectraPro-275
triple grating spectrograph which illuminates an intensified Princeton Instruments
liquid nitrogen cooled 1024 -EUV CCD detector. The probe light is gated
by a PG-200 Programmable Pulse Generator using an MCP intensifier as the
gate source. The intensifier is red/blue enhanced to improve performance
in the UV and red region of the spectrum. The CCD is an array of pixels
(1024x256) with the wavelength being dispersed across the 1024 pixels. In
order to optimize the speed of data collection the software allows for the
'binning together of many pixels to form one 'super' pixel which is read.
All spectra are normally recorded using a single beam configuration, however,
the system is equipped with bifurcated fiber optics cables which allows
for both reference and sample to be collected at once. The standard procedure
for constructing a transient spectrum is first to accumulate a reference
spectrum taken with the laser off, (Io). Finally the sample is collected
at a predetermined time after the laser has been fired and a duration equal
to the reference with laser on, (It). |
 |
Three
dimensional spectra consisting of Change in Absorbance vs. time vs. wavelength
are constructed by obtaining transient difference spectra at many different
time delays. When using a bifurcated cable the Io and It can be collected
at the same time. In this case the probe beam is split into two beams and
the differing amount of light must be accounted for by equating a region
of the spectra with no absorbance changes. |
| Time
resolved spectrum of FeIII(TMPS) + NO as recorded by CCD camera. |
|
|
|
| High
Pressure Measurements |
 |
 |
| Back
to the Top |