|
Safety
The magnetic field extends roughly 5 feet from the center of the magnet.
Do not bring any cell phones, laptops, floppy disks, credit cards, or
mechanical watches near the magnets as they will be permanently damaged.
Do not bring metal objects near the magnets as they can damage the instrument
or cause personal injury. Please refer to the NMR
Safety Guidelines for complete and detailed information.
Logging In
After entering your login name and password, you will be asked for your
account number. This is the charge number you use in the stock room.
Starting the VNMR Software
To start the VNMR software click on the VNMR icon (the one that looks
like an NMR spectrum) on the menu bar at the bottom of the screen.
Loading the Parameters
In your folder there should be some parameters for 1H and
13C (e.g. h1parmnew.200, c13parmcdcl3.200). There are more
parameter files in the vnmr1 folder - click on File, Set
Directory, Parent, select vnmr1 from the list and
then click on Change. To load the parameters that you need, click
on Main Menu, File, select the parameter file from the
list and then click on Load. Be sure to load the parameter file for
the correct nucleus and solvent. If you want to load the shim values,
type load='y'. If you need to change the solvent, type solvent='solvent
name'. Type su to set up your changes.
Loading the Sample
Insert your tube into the spinner and check the tube height using the depth
gauge. Click on Acqi to bring up the acquisition window. Click
on LOCK to open the lock window and turn the spinner off. Click
eject to eject the standard and place it in the holder on the
wall. Place your sample in the barrel on the air cushion and then click
insert. Turn the spinner back on.
Locking and Shimming your Sample
First make sure the sample is spinning; the spin counter should be green
and showing a rate of 20 Hz. Now make sure that LOCK is turned on. This
locks onto the deuterium signal of your solvent to stabilize the magnetic
field. The lock indicator should be green, and the signal should look
like a square wave:
If the lock level is very low or you do not see a square wave:
- Raise the lock power and lock gain values to the maximum by
right clicking on the buttons. If this doesn't work, go to step 2.
- With LOCK turned off, make positive or negative adjustments (right
or left mouse clicks, respectively) to Z0 until you see the
square wave. Turn LOCK back on. The lock signal and square wave
may reach a maximum. Lower the lock power and lock gain
until the square wave becomes visible again; the lock power
should be about 10 less than the lock gain. Always keep the
lock level above 30 or you will lose the lock signal.
The square wave should be positive (like the picture above). If it is
negative, adjust the lock phase until it is positive. Optimize
the lock phase by making positive or negative adjustments until
you find the maximum lock level.
To shim on your sample, click on SHIM to bring up the shim menu. The
coarse shims come up by default, but the only shims you will need
to adjust are the fine shims: Z1, Z2, Z3, Z4, and Z5.
To get to the fine shims, click on the shim button under the lock
level bar.
- Make positive or negative adjustments to Z1 until you reach a
maximum lock level.
- Adjust Z2 until you reach a maximum lock level (as in step 1).
- Repeat steps 1 and 2 until there is no increase in the lock level.
- Cycle through the shims (Z1-Z5) in this fashion until there is
no improvement in the lock level.
- If for some reason you exceed the range of the Z1 or Z2
shims, move them back to the middle and adjust the coarse shims
Z1C or Z2C to bring the lock level back up. Use Z1
or Z2 for fine tuning.
If you need more information, please refer to the User's Handouts titled
"Locking" and "Shimming."
Setting Up Your Experiment
The default parameters are for the MERCURY Vx 200 MHz spectrometer and are good for most
experiments on that instrument. The only ones you
might want to change are sw, pw, d1 and nt.
Acquisition Parameters
| Parameter |
Default |
Description |
| sfrq |
199.972 |
Spectrometer Frequency in MHz. This is a NOT an adjustable parameter.
|
| tn |
H1 |
Transmitter Nucleus. It is set to 1H and is not adjustable.
|
| sw |
3000.3 |
Spectral Width in Hz.
The default value of 3000.3 will give you a 15 ppm window.
(3000.3 Hz/199.972 MHz = 15 ppm). The window must be large enough
to accommodate all possible signals from your sample. If you have a strangely placed
peak or a peak which cannot be phased, the peak is most likely outside of your sw
and has folded into the opposite end of the spectrum. For example, if your spectral window
is set from 10 to -5 ppm, a signal at 12 ppm would fold and show up at -3 ppm.
Conservative settings for the spectral window are an sw of 15 ppm spanning from 13 to -2 ppm.
If you need to make adjustments to your spectral window, follow the procedure described
in the User's Handout titled "Setting the Spectral Window." |
| at |
2 |
Acquisition Time in s. This is the length of time that the FID
is collected for each transient (See the figure at the end of this section). In practice this parameter
is best adjusted indirectly by changing sw or np and letting the computer
calculate at accordingly. |
| np |
12002 |
Number of Points. For digital storage, a FID is represented by data points.
np must be at least twice sw. A value of 4*sw is recommended
for good digital resolution. |
| fb |
1600 |
Filter Bandwidth. This parameter is automatically adjusted by the computer when
sw is changed. |
| bs |
4 |
Block Size. bs represents
the minimum number of transients that need to be acquired before you
can perform a weighted Fourier transform of your spectrum in the foreground
while acquiring more transients in the background. The instrument only
saves data in completed blocks. In practice, a bs of 4
is advisable to assure that at most only three transients are lost when acquisition is
stopped or aborted. |
| ss |
0 |
Steady state transients. ss is the number of times the
instrument will perform the desired pulse sequence without data collection
prior to carrying out the acquisition of real data (sometimes known as dummy scans).
This is done to magnetically prepare the spins in the sample prior to acquiring
an NOE measurement. ss is usually set to zero unless you are doing an
NOE experiment. |
| tpwr |
52 |
Transmitter Power in dB. This is the pulse power. It can be adjusted but MUST NOT EXCEED 60.
|
| pw |
18 |
Pulse Width in μs. This pulse is applied just prior to
acquisition (See the figure at the end of this section). pw = 18 is approximately
a 60° pulse for proton. For some experiments
it is necessary to have a well-calibrated 90° or 180°
pulse (e.g. T1 measurement by inversion recovery). If you need to calibrate the pulse
time for your experiment, refer to the User's Handout titled
"90° Pulse Calibration" for instructions.
When the instrument cannot adjust the receiver gain because your signal is too strong,
a message about ACD overflow will appear on the screen. To fix this, lower the
pw. The coils detect the projection of the net magnetization of your sample in the x,y-plane.
A shorter pw will give a smaller angle, and thus reduce the projected magnetization detected
by the coils. Another advantage of a smaller pulse angle is shorter relaxation time to the pre-pulse
magnetization. |
| p1 |
0 |
First Pulse Length in μs. In multiple pulse experiments (e.g.
inversion recovery), this is the length of the first pulse. |
| d1 |
1 |
First Relaxation Delay in s. This is the delay time before the first pulse in both
single and multi-pulse experiments (see the figure at the end of this section).
The relaxation delay is very important.
If it is too short, the pre-pulse magnetization will not be
restored prior to the next pulse and the resulting spectrum will not provide
quantitative data. If the integrals
for your spectrum do not make sense, increase d1 and re-acquire.
The optimum d1 is at least 5 times your longest T1 time. |
| d2 |
0 |
Second Relaxation Delay in s. This is the delay time between pulses in a standard two-pulse sequence.
|
| tof |
14.8 |
Transmitter Offset Frequency in Hz. tof is the shift in frequency
required to put the carrier frequency in the middle of sw.
Different solvents will have different values of tof due to their
deuterium chemical shifts. To optimize tof,
follow the procedure in the User's Handout titled "Setting the Spectral Window."
|
| nt |
4 |
Number of Transients. This parameter sets the number of FIDs to be acquired in
the experiment. It should be set as a multiple of 4; 40 is good for most research samples.
The more scans you take the better the
signal to noise ratio (S/N) will be. S/N increases as the square
root of the number of transients. This means that if you wish to double your S/N
you must collect 4 times as many transients. |
| ct |
0 |
Completed Transients. This shows how many scans have been completed during your experiment. It is
updated by the computer during acquisition. |
| gain |
Not used |
Receiver Gain. This parameter represents the amount of amplification of the signal.
gain='n', or gain not used, is actually auto gain and the computer controls
the level. In some experiments, such as arrayed experiments, auto gain cannot be used.
gain='y' uses a fixed value. |
Sample Parameters:
| Parameter |
Default |
Description |
| solvent |
CDCl3 |
To change the solvent, type solvent='solvent name' using the appropriate name or formula for your solvent. |
| file |
h1parmnew1.200.fid |
This tells you the filename of the current experiment.
|
Decoupling Parameters:
Please refer to the User's Handout titled "Decoupling."
Processing Parameters:
| Parameter |
Default |
Description |
| lb |
0.10 |
Line broadening. Sets artificial line broadening and exponential weighting along the directly
detected dimension. |
| fn |
32768 |
Fourier number. Selects the Fourier number for the Fourier Transform along the
directly detected dimension. If fn < np, only fn
points are transformed. If fn > np, fn minus
np zeros are added to the data table (zero filling). |
Acquiring a Spectrum
Before acquiring data, you MUST lock and shim on your sample.
- After optimizing the parameters, it is always best to setup your changes
by typing su. To run your experiment and automatically perform a weighted
Fourier transform after completing nt FIDS, type ga (go acquire).
If you do not want the console to automatically perform the transform
type go instead.
- There are two ways to abort the acquisition prior to completion
of nt. Typing sa will stop the acquisition but allows you to continue
by typing ra (resume acquisition). sa/ra can be thought of as
a pause function. Clicking on Abort Acq will abort
the experiment such that it cannot be restarted.
Transforming Your FID
After completion of a block of transients, you can Fourier transform the acquired FIDs
by typing wft (weight and Fourier transform). This can be done many times during
acquisition. After execution of the command your spectrum will be displayed on the screen.
To get the full window, type f.
Processing Your Spectrum
The middle mouse button controls the vertical scale (vs). To increase
the vs hold down the button and drag the mouse up. wp
controls the width of the plot, and sp controls the start
of the plot. The left button controls the left cursor and the right
button controls the right cursor. wc controls the width
of the chart (or window). The ds command will display
the spectrum.
Phasing Your Spectrum
To phase the spectrum type aph to automatically phase the spectrum.
To manually phase the spectrum click on Phase and then
click on the peak that needs to be phased. The left mouse button controls
the course phase and the right mouse button controls the fine phase.
Creating a Scale and Changing the Axis Units
To place a scale under the spectrum, type dscale for display
scale or click on the Dscale button.
To change the units of the scale type axis='h' for Hz and
axis='p' for ppm.
Referencing the Solvent
To reference the solvent, zoom in on the solvent peak by surrounding it
with the two cursors. Then click expand.
Place a single cursor on the solvent peak, and type nl
for nearest line. This will center the line in the middle
of the peak. If the chemical shift is incorrect click on Ref or
type rl(#.##p).
(e.g. For CDCl3 you would type rl(7.27p)).
Integration
There are three integration modes: part, full, and none. By clicking on
Integral you can change the mode that you are in. The "part"
mode lets you break up the integral. The "full" mode gives
you a single integral.
- Breaking up the integral. First click on Part Integral.
When the integral comes up, type cz
to clear the memory (clear zeros). Click on Lvl/Tlt to interactively adjust
the integral line with the mouse. Then click on Resets.
To break up the integral, left click on the left side of a peak
then the right side. Repeat this until you integrate all of the peaks that
you want. The ins command is used to normalize the areas. By
typing ins=100 all of the areas on the screen will add up to 100.
If you want absolute areas, type ins=1 and insref=l. (Note:
you can use insref=le-5 to move the decimal point.)
- There are two ways to display the integral areas:
- Typing dli (display list of integrals) displays
the areas in the text window. To print out the list of integral areas
type printon dli printoff.
- To display the areas on the spectrum, first adjust the vertical position by
typing vp=12, then type dpirn for normalized areas or
dpir for absolute areas.
Peak Picking
Press the Th button and use the left mouse button to
raise or lower the yellow line. Type dpf (display peak frequencies) to
label the peaks. All peaks above the line will be labeled.
Creating a Text Label
To create a text label type text('your label').
Configuring the Plotter
On the 200 MHz NMR, there are two printers that you can use. For standard size paper use the printer
next to the computer (LaserPlot6L). For the large 11x17 paper choose the printer in the 3rd floor
NMR room (LaserPlot800011x17). To select the printer, click on Main Menu,
More, Configure, Select Plotter. This changes the selected printer
to the next option. The current printer will be displayed in the text window above the command line.
Repeat this entire process until you've selected the printer that you want.
Printing your Spectrum
To print out the spectrum, type the appropriate print commands ending with page.
| pl |
Plots the spectrum |
| pscale |
Prints the scale |
| pir |
Prints the integral areas |
| ppf |
Prints the peak frequencies |
| pap |
Prints all of the parameters |
| ptext |
Prints the text label |
| page |
Gives you a new page |
Saving your FID
You can only save FIDs in your own folder. If you are in a different
folder, return to your own folder by clicking on Main Menu,
File, Set Directory, Home. To save your FID,
type svf (save FID). It will ask what you want to call it.
Type the filename followed by the following suffixes:
| .200 |
200 MHz spectrometer |
| .qnp |
400 MHz spectrometer |
| .idp |
500 MHz indirect detection probe |
| .swp |
500 MHz switchable |
| .bbp |
500 MHz broad band probe |
Joining Different Experiments
To join a different experiment, type jexp#, where # is the experiment
number.
Logging Out
When you are done, eject your sample and insert the ethyl benzene standard. After locking and shimming,
take 4 scans to check the resolution of the TMS peak. Zoom into the TMS region, the peak should be a
singlet. Place the cursor
on the TMS peak and type nl then dres (display resolution).
If the resolution is below 1 Hz you can log out.
If not, you must reshim until the linewidth is < 1 Hz.
To log out, click on Main Menu, More and Exit VNMR.
This will get you out of the VNMR software.
To log off the system, click on the EXIT button
on the menu bar at the bottom of the screen. Be sure to sign the log book
and appropriate charge sheet.
|